This is not the current EPA website. To navigate to the current EPA website, please go to www.epa.gov. This website is historical material reflecting the EPA website as it existed on January 19, 2021. This website is no longer updated and links to external websites and some internal pages may not work. More information »

Memorandum Regarding Use of Hydrochloric Acid (HCl) in Digests for Inductively Coupled Plasma-Mass Spectrometry (ICP-MS) Analysis

This memorandum clarifies which version of Method 3050 to use based on the analysis technique. If inductively coupled plasma – atomic emission spectroscopy (ICP-AES) is used, HCl should be used in the acid digestion, following Method 3050A. If ICP-MS is used, no HCl should be used due to matrix interference, following Method 3050B. The memorandum also discusses the detection of silver and antimony by ICP-MS.